Scan Abutment WP, Cat. No. 30-70106 - Product Usage: used to take an impression with an intra oral scanner or optical scanner to provide...
FDA Recall Notice
Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-0018-2022.
Some devices are outside tolerances and showed a maximal deviation of 2 degrees in the orientation between the peek head and titanium base, and an incongruence between the size engraved on the peek has been identified. Deviations may delay the final restoration.
Corrective Action (per FDA)
Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-0018-2022.
Recall terminated by FDA.
✅ What you should do
- Stop using the product if you own it.
- Check the model number, lot code, or sell-by date against the recall notice above.
- Contact Paltop Advanced Dental Solutions Ltd or the retailer where you bought it for a refund, replacement, or repair.
- For the most current official instructions, visit the FDA recall page.
- If you've been hurt by this product, report the incident to FDA.
Consumer Contact (per FDA)
Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-0018-2022.
Paltop Advanced Dental Solutions Ltd
About the U.S. Food and Drug Administration
The FDA regulates drugs, medical devices, food, cosmetics, and tobacco. Adverse event reports and recall notices are the main public safety signal.
Visit FDA.gov →📣 Report a food, supplement, or cosmetic problem to the FDA
If you had a reaction, found contamination, or experienced a labeling problem with this product, report it to the FDA. The agency uses consumer reports to track emerging safety signals and trigger recalls.
Paltop Advanced Dental Solutions Ltd Recall FAQ
Paltop Advanced Dental Solutions Ltd is the subject of a medical implants safety report: Scan Abutment WP, Cat. No. 30-70106 - Product Usage: used to take an impression with an intra oral scanner or optical scanner to provide.... The notice was published on July 26, 2021 by the U.S. Food and Drug Administration (FDA). Approximately 236 units are potentially affected.