Tritanium TL Steerable Inserter, Cat. No. 48930000, UDI 07613327381900 Product Usage: The TL Steerable Inserter is a Class 1 instrume...
FDA Recall Notice
Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-0207-2019.
The firm received complaints related to the locking mechanism on the steerable inserter not functioning as intended. The firm confirmed that the issue is related to lack of lubrication of the instrument. There are two potential hazards associated with the implant pivoting unexpectedly during implant insertion into the disc space. First, if the implant is initially inserted into the disc space via the impaction method and the implant pivots unexpectedly, the implant potentially may unintentionally contact sensitive anatomy. Additionally, the user may potentially experience difficulty positioning the implant within the disc space due to the lack of rigidity. A third potential hazard is associated with the inserter not unlocking at the articulating pivot head during implant positioning within the disc space. Once the implant is in the prepared disc space and contacts the ventral annulus, the surgeon unlocks the articulating pivot head and positions the implant perpendicular to the spinous process. The surgeon may then elect to lock/unlock the articulating pivot head for desired final positioning of the implant. If the inserter does not unlock during these final implant positioning steps, the user may potentially implant the cage in a suboptimal trajectory/placement.
Corrective Action (per FDA)
Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-0207-2019.
Recall terminated by FDA.
🩺 Talk to your doctor or pharmacist
Product Recall Tracker does not provide medical advice. Any specific instructions about this medication or device come directly from the FDA notice above. Before making any change to your treatment:
- Do not stop, change, or discard your prescription on your own.
- Call your doctor or pharmacist and ask whether this FDA notice applies to your prescription (have the brand name, dosage, and lot number ready if you can find them).
- Follow your clinician's guidance on whether to continue, switch, or return the product.
- If you have experienced a reaction, seek medical care and report it to the FDA via MedWatch.
- For the official FDA notice, visit the FDA recalls page.
Consumer Contact (per FDA)
Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-0207-2019.
Howmedica Osteonics Corp.
About the U.S. Food and Drug Administration
The FDA regulates drugs, medical devices, food, cosmetics, and tobacco. Adverse event reports and recall notices are the main public safety signal.
Visit FDA.gov →📣 Report an adverse event to the FDA (MedWatch)
If you had an adverse reaction, side effect, or device malfunction from this medication or device, report it to the FDA through MedWatch. Talk to your doctor or pharmacist first — do not stop a prescription on your own. Reports help the FDA detect rare or emerging safety signals and can lead to label changes or recalls.
Howmedica Osteonics Corp. Recall FAQ
Howmedica Osteonics Corp. is the subject of a medical devices safety report: Tritanium TL Steerable Inserter, Cat. No. 48930000, UDI 07613327381900 Product Usage: The TL Steerable Inserter is a Class 1 instrume.... The notice was published on September 12, 2018 by the U.S. Food and Drug Administration (FDA). Approximately 58 units are potentially affected.