Abbott Alinity m STI AMP Kit, used with the Alinity m System, product codes: a) REF 09N17-095; b) REF 09N17-090; c) REF 09N17-091;
FDA Recall Notice
Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-0234-2025.
Abbott has identified an increase of incidences regarding Error Code (EC) 9198 (Positive control is non-reactive) while using the Alinity m HR HPV AMP Kit and Alinity m STI AMP Kit. Certain invalidated positive assay controls can be traced to iron leaching into the Alinity m Lysis Solution from the lysis transfer pump in the Alinity m System.
Corrective Action (per FDA)
Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-0234-2025.
Recall ongoing. Follow firm instructions.
🩺 Talk to your doctor or pharmacist
Product Recall Tracker does not provide medical advice. Any specific instructions about this medication or device come directly from the FDA notice above. Before making any change to your treatment:
- Do not stop, change, or discard your prescription on your own.
- Call your doctor or pharmacist and ask whether this FDA notice applies to your prescription (have the brand name, dosage, and lot number ready if you can find them).
- Follow your clinician's guidance on whether to continue, switch, or return the product.
- If you have experienced a reaction, seek medical care and report it to the FDA via MedWatch.
- For the official FDA notice, visit the FDA recalls page.
Consumer Contact (per FDA)
Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-0234-2025.
Abbott Molecular, Inc.
About the U.S. Food and Drug Administration
The FDA regulates drugs, medical devices, food, cosmetics, and tobacco. Adverse event reports and recall notices are the main public safety signal.
Visit FDA.gov →📣 Report an adverse event to the FDA (MedWatch)
If you had an adverse reaction, side effect, or device malfunction from this medication or device, report it to the FDA through MedWatch. Talk to your doctor or pharmacist first — do not stop a prescription on your own. Reports help the FDA detect rare or emerging safety signals and can lead to label changes or recalls.
Abbott Molecular, Inc. Recall FAQ
Abbott Molecular, Inc. is the subject of a medical devices safety report: Abbott Alinity m STI AMP Kit, used with the Alinity m System, product codes: a) REF 09N17-095; b) REF 09N17-090; c) REF 09N17-091;. The notice was published on September 12, 2024 by the U.S. Food and Drug Administration (FDA). Approximately 15,068 units are potentially affected.