Skip to main content
High RiskFDAfda-Z-0436-2013BATTERY HAZARD

220/240V AC, Advanced Perfusion System Platform (APS) The Terumo¿ Advanced Perfusion System 1 is indicated for use for up to 6 hours o...

Units Affected
1,647
Recall Date
November 14, 2012
Issuing Agency
Hazard
Battery Hazard

FDA Recall Notice

Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-0436-2013.

Terumo Cardiovascular System (TCVS) has received reports of a situation where users experienced a total loss of functionality for some System 1 units. The reports indicate that the units went blank and shut down with no sign of power and battery backup did not initiate. The result is all pumps stop, with no safety system functionality, and the battery would not be activated. The user would be

Corrective Action (per FDA)

Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-0436-2013.

Recall terminated by FDA.

🩺 Talk to your doctor or pharmacist

Product Recall Tracker does not provide medical advice. Any specific instructions about this medication or device come directly from the FDA notice above. Before making any change to your treatment:

  1. Do not stop, change, or discard your prescription on your own.
  2. Call your doctor or pharmacist and ask whether this FDA notice applies to your prescription (have the brand name, dosage, and lot number ready if you can find them).
  3. Follow your clinician's guidance on whether to continue, switch, or return the product.
  4. If you have experienced a reaction, seek medical care and report it to the FDA via MedWatch.
  5. For the official FDA notice, visit the FDA recalls page.

Consumer Contact (per FDA)

Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-0436-2013.

Terumo Cardiovascular Systems Corporation

FDA

About the U.S. Food and Drug Administration

The FDA regulates drugs, medical devices, food, cosmetics, and tobacco. Adverse event reports and recall notices are the main public safety signal.

Visit FDA.gov →

📣 Report an adverse event to the FDA (MedWatch)

If you had an adverse reaction, side effect, or device malfunction from this medication or device, report it to the FDA through MedWatch. Talk to your doctor or pharmacist first — do not stop a prescription on your own. Reports help the FDA detect rare or emerging safety signals and can lead to label changes or recalls.

How to report to FDAFile a report at FDA

Terumo Cardiovascular Systems Corporation Recall FAQ

Terumo Cardiovascular Systems Corporation is the subject of a medical devices safety report: 220/240V AC, Advanced Perfusion System Platform (APS) The Terumo¿ Advanced Perfusion System 1 is indicated for use for up to 6 hours o.... The notice was published on November 14, 2012 by the U.S. Food and Drug Administration (FDA). Approximately 1,647 units are potentially affected.