Bio-Modular Humeral Head 48mm x 24mm Ti 6AI4V/Ion Implanted (Orthopedic Implant) Product Usage: The OSS Knee Reamer Sleeves are prima...
FDA Recall Notice
Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-0474-2013.
The items in this lot are missing the ion implantation feature. The surface of the head may be softer and more prone to scratching, which in tum could potentially cause higher wear of the poly bearing. Increased wear could potentially lead to a shorter implant life.
Corrective Action (per FDA)
Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-0474-2013.
Recall terminated by FDA.
✅ What you should do
- Stop using the product if you own it.
- Check the model number, lot code, or sell-by date against the recall notice above.
- Contact Biomet, Inc. or the retailer where you bought it for a refund, replacement, or repair.
- For the most current official instructions, visit the FDA recall page.
- If you've been hurt by this product, report the incident to FDA.
Consumer Contact (per FDA)
Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-0474-2013.
Biomet, Inc.
About the U.S. Food and Drug Administration
The FDA regulates drugs, medical devices, food, cosmetics, and tobacco. Adverse event reports and recall notices are the main public safety signal.
Visit FDA.gov →📣 Report a food, supplement, or cosmetic problem to the FDA
If you had a reaction, found contamination, or experienced a labeling problem with this product, report it to the FDA. The agency uses consumer reports to track emerging safety signals and trigger recalls.
Biomet, Inc. Recall FAQ
Biomet, Inc. is the subject of a medical implants safety report: Bio-Modular Humeral Head 48mm x 24mm Ti 6AI4V/Ion Implanted (Orthopedic Implant) Product Usage: The OSS Knee Reamer Sleeves are prima.... The notice was published on November 1, 2012 by the U.S. Food and Drug Administration (FDA). Approximately 4 units are potentially affected.