Skip to main content
High RiskFDAfda-Z-0574-2020FALL HAZARD

ARCHITECT HAVAB-G Calibrator, REF 6L27-01

Units Affected
83
Recall Date
November 7, 2019
Issuing Agency
Hazard
Fall Hazard

FDA Recall Notice

Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-0574-2020.

Abbott has identified that Calibrator lot 08219BE00 is potentially impacted by an increase in Relative Light Units (RLUs) up to 94%. It has been confirmed that the increase in the calibrator RLU values may cause a decrease in the control values and patient results. Patient values generated using the impacted calibrator lot 08219BE00 may have changed interpretation from reactive to falsely non reactive. Results impacted may present as a (falsely) non reactive result that fall in the range between 0.50 to 0.99 S/CO.

Corrective Action (per FDA)

Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-0574-2020.

Recall terminated by FDA.

🩺 Talk to your doctor or pharmacist

Product Recall Tracker does not provide medical advice. Any specific instructions about this medication or device come directly from the FDA notice above. Before making any change to your treatment:

  1. Do not stop, change, or discard your prescription on your own.
  2. Call your doctor or pharmacist and ask whether this FDA notice applies to your prescription (have the brand name, dosage, and lot number ready if you can find them).
  3. Follow your clinician's guidance on whether to continue, switch, or return the product.
  4. If you have experienced a reaction, seek medical care and report it to the FDA via MedWatch.
  5. For the official FDA notice, visit the FDA recalls page.

Consumer Contact (per FDA)

Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-0574-2020.

Abbott Gmbh & Co. KG

FDA

About the U.S. Food and Drug Administration

The FDA regulates drugs, medical devices, food, cosmetics, and tobacco. Adverse event reports and recall notices are the main public safety signal.

Visit FDA.gov →

📣 Report an adverse event to the FDA (MedWatch)

If you had an adverse reaction, side effect, or device malfunction from this medication or device, report it to the FDA through MedWatch. Talk to your doctor or pharmacist first — do not stop a prescription on your own. Reports help the FDA detect rare or emerging safety signals and can lead to label changes or recalls.

How to report to FDAFile a report at FDA

Abbott Gmbh & Co. Kg Recall FAQ

Abbott Gmbh & Co. Kg is the subject of a medical devices safety report: ARCHITECT HAVAB-G Calibrator, REF 6L27-01. The notice was published on November 7, 2019 by the U.S. Food and Drug Administration (FDA). Approximately 83 units are potentially affected.