Skip to main content
CriticalFDAfda-Z-0662-2025OTHER

Exacta External Drainage and Monitoring System: 27581 KIT 27581 EXACTA W/EDM VCATH 35CM 27636 EDM 27636 EXACTA W/BRAIDED CORD LOCK 276...

⚠ Critical FDA Safety Notice — Talk to Your Doctor

The FDA has flagged this as a serious safety concern (other). Do not stop, change, or discard a prescribed medication or medical device on your own. Contact your doctor or pharmacist immediately to ask whether this notice applies to your prescription and what to do next. Any specific guidance below comes directly from the FDA — Product Recall Tracker does not provide medical advice.

Units Affected
258,367
Recall Date
November 1, 2024
Issuing Agency
Hazard
Other

FDA Recall Notice

Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-0662-2025.

Due to complaints received reporting cracks and/or leaks of the stopcocks associated with the external drainage and monitoring system (EDMS).

Corrective Action (per FDA)

Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-0662-2025.

Recall ongoing. Follow firm instructions.

🩺 Talk to your doctor or pharmacist

Product Recall Tracker does not provide medical advice. Any specific instructions about this medication or device come directly from the FDA notice above. Before making any change to your treatment:

  1. Do not stop, change, or discard your prescription on your own.
  2. Call your doctor or pharmacist and ask whether this FDA notice applies to your prescription (have the brand name, dosage, and lot number ready if you can find them).
  3. Follow your clinician's guidance on whether to continue, switch, or return the product.
  4. If you have experienced a reaction, seek medical care and report it to the FDA via MedWatch.
  5. For the official FDA notice, visit the FDA recalls page.

Consumer Contact (per FDA)

Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-0662-2025.

Medtronic Neurosurgery

FDA

About the U.S. Food and Drug Administration

The FDA regulates drugs, medical devices, food, cosmetics, and tobacco. Adverse event reports and recall notices are the main public safety signal.

Visit FDA.gov →

📣 Report an adverse event to the FDA (MedWatch)

If you had an adverse reaction, side effect, or device malfunction from this medication or device, report it to the FDA through MedWatch. Talk to your doctor or pharmacist first — do not stop a prescription on your own. Reports help the FDA detect rare or emerging safety signals and can lead to label changes or recalls.

How to report to FDAFile a report at FDA

Medtronic Neurosurgery Recall FAQ

Medtronic Neurosurgery is the subject of a medical devices safety report: Exacta External Drainage and Monitoring System: 27581 KIT 27581 EXACTA W/EDM VCATH 35CM 27636 EDM 27636 EXACTA W/BRAIDED CORD LOCK 276.... The notice was published on November 1, 2024 by the U.S. Food and Drug Administration (FDA). Approximately 258,367 units are potentially affected.