High RiskFDAfda-Z-0753-2025SOFTWARE DEFECT

The Merlin PCS 3650 programmer Model 3330 The Merlin PCS programmer model 3650 is a portable, dedicated programming system which oper...

Units Affected
7,295
Recall Date
November 21, 2024
Issuing Agency
Hazard
Software Defect

FDA Recall Notice

Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-0753-2025.

Due to a programmer software anomaly, during finalizing of the leadless pacemaker system, a specific sequence of programmer actions combined with a loss of telemetry during a small window (<2 seconds) may cause the finalization step to fail. If this occurs, the Programmer displays a loss of telemetry detected window, and there is no pacing.

Corrective Action (per FDA)

Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-0753-2025.

Recall ongoing. Follow firm instructions.

✅ What you should do

  1. Stop using the product if you own it.
  2. Check the model number, lot code, or sell-by date against the recall notice above.
  3. Contact Abbott Medical or the retailer where you bought it for a refund, replacement, or repair.
  4. For the most current official instructions, visit the FDA recall page.
  5. If you've been hurt by this product, report the incident to FDA.

Consumer Contact (per FDA)

Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-0753-2025.

Abbott Medical

FDA

About the U.S. Food and Drug Administration

The FDA regulates drugs, medical devices, food, cosmetics, and tobacco. Adverse event reports and recall notices are the main public safety signal.

Visit FDA.gov →

📣 Report a food, supplement, or cosmetic problem to the FDA

If you had a reaction, found contamination, or experienced a labeling problem with this product, report it to the FDA. The agency uses consumer reports to track emerging safety signals and trigger recalls.

How to report to FDAFile a report at FDA

Abbott Medical Recall FAQ

Abbott Medical is the subject of a medical implants safety report: The Merlin PCS 3650 programmer Model 3330 The Merlin PCS programmer model 3650 is a portable, dedicated programming system which oper.... The notice was published on November 21, 2024 by the U.S. Food and Drug Administration (FDA). Approximately 7,295 units are potentially affected.