High RiskFDAfda-Z-0778-2021OTHER

Hot AXIOS Stent and Electrocautery - Enhanced Delivery System 20mm x10mm (OUS) UPN: M00553560 - Product Usage: The stent is intended for ...

Units Affected
78
Recall Date
December 1, 2020
Issuing Agency
Hazard
Other

FDA Recall Notice

Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-0778-2021.

The rotating luer fitting at the distal end of the delivery system handle may become detached from the nose of the delivery during device preparation, stent delivery or use, resulting in delay or procedure or risk of peritonitis developing from leakage of enteric, pseudocyst, or biliary contents into the peritoneal cavity

Corrective Action (per FDA)

Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-0778-2021.

Recall terminated by FDA.

✅ What you should do

  1. Stop using the product if you own it.
  2. Check the model number, lot code, or sell-by date against the recall notice above.
  3. Contact Boston Scientific Corporation or the retailer where you bought it for a refund, replacement, or repair.
  4. For the most current official instructions, visit the FDA recall page.
  5. If you've been hurt by this product, report the incident to FDA.

Consumer Contact (per FDA)

Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-0778-2021.

Boston Scientific Corporation

FDA

About the U.S. Food and Drug Administration

The FDA regulates drugs, medical devices, food, cosmetics, and tobacco. Adverse event reports and recall notices are the main public safety signal.

Visit FDA.gov →

📣 Report a food, supplement, or cosmetic problem to the FDA

If you had a reaction, found contamination, or experienced a labeling problem with this product, report it to the FDA. The agency uses consumer reports to track emerging safety signals and trigger recalls.

How to report to FDAFile a report at FDA

Boston Scientific Corporation Recall FAQ

Boston Scientific Corporation is the subject of a medical implants safety report: Hot AXIOS Stent and Electrocautery - Enhanced Delivery System 20mm x10mm (OUS) UPN: M00553560 - Product Usage: The stent is intended for .... The notice was published on December 1, 2020 by the U.S. Food and Drug Administration (FDA). Approximately 78 units are potentially affected.