Skip to main content
High RiskFDAfda-Z-0940-2026OTHER

MADSEN ACCUSCREEN TEOAE/DPOAE/ABR Probe (P/N 8-69-41100)

Brand
Units Affected
0
Recall Date
November 6, 2025
Issuing Agency
Hazard
Other

FDA Recall Notice

Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-0940-2026.

Following a component change in the DPOAE probe (speaker replacement), technical distortions may occur at high stimulus levels combined with unusually low OAE amplitude acceptance thresholds. These distortions can mimic true DPOAE responses and may result in a false PASS outcome.

Corrective Action (per FDA)

Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-0940-2026.

Recall ongoing. Follow firm instructions.

🩺 Talk to your doctor or pharmacist

Product Recall Tracker does not provide medical advice. Any specific instructions about this medication or device come directly from the FDA notice above. Before making any change to your treatment:

  1. Do not stop, change, or discard your prescription on your own.
  2. Call your doctor or pharmacist and ask whether this FDA notice applies to your prescription (have the brand name, dosage, and lot number ready if you can find them).
  3. Follow your clinician's guidance on whether to continue, switch, or return the product.
  4. If you have experienced a reaction, seek medical care and report it to the FDA via MedWatch.
  5. For the official FDA notice, visit the FDA recalls page.

Consumer Contact (per FDA)

Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-0940-2026.

PATH

FDA

About the U.S. Food and Drug Administration

The FDA regulates drugs, medical devices, food, cosmetics, and tobacco. Adverse event reports and recall notices are the main public safety signal.

Visit FDA.gov →

📣 Report an adverse event to the FDA (MedWatch)

If you had an adverse reaction, side effect, or device malfunction from this medication or device, report it to the FDA through MedWatch. Talk to your doctor or pharmacist first — do not stop a prescription on your own. Reports help the FDA detect rare or emerging safety signals and can lead to label changes or recalls.

How to report to FDAFile a report at FDA

Path Recall FAQ

Path is the subject of a medical devices safety report: MADSEN ACCUSCREEN TEOAE/DPOAE/ABR Probe (P/N 8-69-41100). The notice was published on November 6, 2025 by the U.S. Food and Drug Administration (FDA). Approximately 0 units are potentially affected.