Skip to main content
CriticalFDAfda-Z-1290-2015OTHER

Diamondback 360 Coronary Orbital Atherectomy System, Model Number: DBEC-125. The system is indicated to facilitate stent delivery in p...

⚠ Critical Alert — Stop Using Immediately

This product has been flagged with severe risks (other). Stop using it now and contact the brand or FDA for a refund, repair, or replacement.

Units Affected
30
Recall Date
March 4, 2015
Issuing Agency
Hazard
Other

FDA Recall Notice

Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-1290-2015.

The OAD saline sheath outer layer may flake off from the catheter due to friction between the outer material layer and corresponding materials it may come in contact with during clinical use. If the material is large enough, it has the potential to cause a flow limiting embolism.

Corrective Action (per FDA)

Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-1290-2015.

Recall terminated by FDA.

✅ What you should do

  1. Stop using the product if you own it.
  2. Check the model number, lot code, or sell-by date against the recall notice above.
  3. Contact Cardiovascular Systems, Inc. or the retailer where you bought it for a refund, replacement, or repair.
  4. For the most current official instructions, visit the FDA recall page.
  5. If you've been hurt by this product, report the incident to FDA.

Consumer Contact (per FDA)

Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-1290-2015.

Cardiovascular Systems, Inc.

FDA

About the U.S. Food and Drug Administration

The FDA regulates drugs, medical devices, food, cosmetics, and tobacco. Adverse event reports and recall notices are the main public safety signal.

Visit FDA.gov →

📣 Report a food, supplement, or cosmetic problem to the FDA

If you had a reaction, found contamination, or experienced a labeling problem with this product, report it to the FDA. The agency uses consumer reports to track emerging safety signals and trigger recalls.

How to report to FDAFile a report at FDA

Cardiovascular Systems, Inc. Recall FAQ

Cardiovascular Systems, Inc. is the subject of a medical implants safety report: Diamondback 360 Coronary Orbital Atherectomy System, Model Number: DBEC-125. The system is indicated to facilitate stent delivery in p.... The notice was published on March 4, 2015 by the U.S. Food and Drug Administration (FDA). Approximately 30 units are potentially affected.