DxI 9000 Access Immunoassay Analyzer, Catalog Number C11137

Units Affected
245
Recall Date
November 13, 2024
Issuing Agency
Hazard
Other

FDA Recall Notice

Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-1326-2025.

The DxI 9000 Access Immunoassay Analyzer wash wheel mixer may not spin freely within the idler pulley housing. When the problem occurs, the analyzer may generate a system event for a Wash Wheel Dispense Arm Spin Mixer Error that is displayed on the navigation bar in the User Interface. The analyzer enters the red system status and will generate a SYS flag. The tests that are in progress may be cancelled and may cause a delay in reporting patient results

Corrective Action (per FDA)

Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-1326-2025.

Recall ongoing. Follow firm instructions.

🩺 Talk to your doctor or pharmacist

Product Recall Tracker does not provide medical advice. Any specific instructions about this medication or device come directly from the FDA notice above. Before making any change to your treatment:

  1. Do not stop, change, or discard your prescription on your own.
  2. Call your doctor or pharmacist and ask whether this FDA notice applies to your prescription (have the brand name, dosage, and lot number ready if you can find them).
  3. Follow your clinician's guidance on whether to continue, switch, or return the product.
  4. If you have experienced a reaction, seek medical care and report it to the FDA via MedWatch.
  5. For the official FDA notice, visit the FDA recalls page.

Consumer Contact (per FDA)

Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-1326-2025.

Beckman Coulter, Inc.

FDA

About the U.S. Food and Drug Administration

The FDA regulates drugs, medical devices, food, cosmetics, and tobacco. Adverse event reports and recall notices are the main public safety signal.

Visit FDA.gov →

📣 Report an adverse event to the FDA (MedWatch)

If you had an adverse reaction, side effect, or device malfunction from this medication or device, report it to the FDA through MedWatch. Talk to your doctor or pharmacist first — do not stop a prescription on your own. Reports help the FDA detect rare or emerging safety signals and can lead to label changes or recalls.

How to report to FDAFile a report at FDA

Beckman Coulter, Inc. Recall FAQ

Beckman Coulter, Inc. is the subject of a diagnostic devices safety report: DxI 9000 Access Immunoassay Analyzer, Catalog Number C11137. The notice was published on November 13, 2024 by the U.S. Food and Drug Administration (FDA). Approximately 245 units are potentially affected.