Skip to main content
High RiskFDAfda-Z-1419-2022LACERATION

ATTUNE Measured Sizing and Rotation Guide-to size the femur and align rotation of the bone cuts in preparation for the femoral implant P...

Units Affected
872
Recall Date
June 29, 2022
Issuing Agency
Hazard
Laceration

FDA Recall Notice

Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-1419-2022.

Manufactured with a Galvanized High Carbon Steel spring instead of the correct 302 Stainless steel spring. has the potential to oxidize and corrode in the ATTUNE Measured Sizing & Rotation Guide, resulting in surgical delay, debris in surgical field, Osteolysis, Poor Joint Mechanics, Soft Tissue Irritation, Pain

Corrective Action (per FDA)

Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-1419-2022.

Recall ongoing. Follow firm instructions.

✅ What you should do

  1. Stop using the product if you own it.
  2. Check the model number, lot code, or sell-by date against the recall notice above.
  3. Contact Depuy Orthopaedics, Inc. or the retailer where you bought it for a refund, replacement, or repair.
  4. For the most current official instructions, visit the FDA recall page.
  5. If you've been hurt by this product, report the incident to FDA.

Consumer Contact (per FDA)

Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-1419-2022.

DePuy Orthopaedics, Inc.

FDA

About the U.S. Food and Drug Administration

The FDA regulates drugs, medical devices, food, cosmetics, and tobacco. Adverse event reports and recall notices are the main public safety signal.

Visit FDA.gov →

📣 Report a food, supplement, or cosmetic problem to the FDA

If you had a reaction, found contamination, or experienced a labeling problem with this product, report it to the FDA. The agency uses consumer reports to track emerging safety signals and trigger recalls.

How to report to FDAFile a report at FDA

Depuy Orthopaedics, Inc. Recall FAQ

Depuy Orthopaedics, Inc. is the subject of a medical implants safety report: ATTUNE Measured Sizing and Rotation Guide-to size the femur and align rotation of the bone cuts in preparation for the femoral implant P.... The notice was published on June 29, 2022 by the U.S. Food and Drug Administration (FDA). Approximately 872 units are potentially affected.