Skip to main content
High RiskFDAfda-Z-1527-2018OTHER

EEA" Hemorrhoid and Prolapse Stapler Set with DST Series" Technology 3.5MM, Product Number HEM3335, GTIN Number 1 - 10884521047518, GTIN ...

Units Affected
5,547
Recall Date
April 9, 2018
Issuing Agency
Hazard
Other

FDA Recall Notice

Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-1527-2018.

There is a potential for improper welding of the yellow staple guide to the instrument. Use of a device with an improperly welded staple guide may result in improper staple formation potentially leading to bleeding or anastomotic leak.

Corrective Action (per FDA)

Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-1527-2018.

Recall terminated by FDA.

✅ What you should do

  1. Stop using the product if you own it.
  2. Check the model number, lot code, or sell-by date against the recall notice above.
  3. Contact Covidien Llc or the retailer where you bought it for a refund, replacement, or repair.
  4. For the most current official instructions, visit the FDA recall page.
  5. If you've been hurt by this product, report the incident to FDA.

Consumer Contact (per FDA)

Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-1527-2018.

Covidien LLC

FDA

About the U.S. Food and Drug Administration

The FDA regulates drugs, medical devices, food, cosmetics, and tobacco. Adverse event reports and recall notices are the main public safety signal.

Visit FDA.gov →

📣 Report a food, supplement, or cosmetic problem to the FDA

If you had a reaction, found contamination, or experienced a labeling problem with this product, report it to the FDA. The agency uses consumer reports to track emerging safety signals and trigger recalls.

How to report to FDAFile a report at FDA

Covidien Llc Recall FAQ

Covidien Llc is the subject of a medical implants safety report: EEA" Hemorrhoid and Prolapse Stapler Set with DST Series" Technology 3.5MM, Product Number HEM3335, GTIN Number 1 - 10884521047518, GTIN .... The notice was published on April 9, 2018 by the U.S. Food and Drug Administration (FDA). Approximately 5,547 units are potentially affected.