MiniMed 670G Insulin Pump: Models: PUMP MMT-1740KX 670G V6.3 SF MG CLIN;¿ PUMP MMT-1740KX 670G V6.3 CLIN 020 MG;¿ PUMP MMT-1741KFX 67...
FDA Recall Notice
Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-1772-2021.
Due to a software design issue, under certain conditions, a software fault is detected when a large bolus delivery at quick bolus speed completes. if the user is not aware of the amount of active insulin and delivers an additional bolus, there is a risk of insulin over delivery.
Corrective Action (per FDA)
Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-1772-2021.
Recall ongoing. Follow firm instructions.
🩺 Talk to your doctor or pharmacist
Product Recall Tracker does not provide medical advice. Any specific instructions about this medication or device come directly from the FDA notice above. Before making any change to your treatment:
- Do not stop, change, or discard your prescription on your own.
- Call your doctor or pharmacist and ask whether this FDA notice applies to your prescription (have the brand name, dosage, and lot number ready if you can find them).
- Follow your clinician's guidance on whether to continue, switch, or return the product.
- If you have experienced a reaction, seek medical care and report it to the FDA via MedWatch.
- For the official FDA notice, visit the FDA recalls page.
Consumer Contact (per FDA)
Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-1772-2021.
Medtronic Minimed
About the U.S. Food and Drug Administration
The FDA regulates drugs, medical devices, food, cosmetics, and tobacco. Adverse event reports and recall notices are the main public safety signal.
Visit FDA.gov →📣 Report an adverse event to the FDA (MedWatch)
If you had an adverse reaction, side effect, or device malfunction from this medication or device, report it to the FDA through MedWatch. Talk to your doctor or pharmacist first — do not stop a prescription on your own. Reports help the FDA detect rare or emerging safety signals and can lead to label changes or recalls.
Medtronic Minimed Recall FAQ
Medtronic Minimed is the subject of a diabetes devices safety report: MiniMed 670G Insulin Pump: Models: PUMP MMT-1740KX 670G V6.3 SF MG CLIN;¿ PUMP MMT-1740KX 670G V6.3 CLIN 020 MG;¿ PUMP MMT-1741KFX 67.... The notice was published on March 15, 2021 by the U.S. Food and Drug Administration (FDA). Approximately 1,266 units are potentially affected.