High RiskFDAfda-Z-1789-2014BATTERY HAZARD

Philips/Children's Medical Ventures Smart Monitor 2 Professional Series infant apnea monitor.

Units Affected
112
Recall Date
April 29, 2014
Issuing Agency
Hazard
Battery Hazard

FDA Recall Notice

Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-1789-2014.

The battery pack wire harness is improperly assembled rendering the device inoperable. The wiring issue causes an error code to be displayed and results in the device sounding an alarm repeatedly at power up. The presence of the error code and the devices alarm prevents the device from functioning until the issue is addressed.

Corrective Action (per FDA)

Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-1789-2014.

Recall terminated by FDA.

🩺 Talk to your doctor or pharmacist

Product Recall Tracker does not provide medical advice. Any specific instructions about this medication or device come directly from the FDA notice above. Before making any change to your treatment:

  1. Do not stop, change, or discard your prescription on your own.
  2. Call your doctor or pharmacist and ask whether this FDA notice applies to your prescription (have the brand name, dosage, and lot number ready if you can find them).
  3. Follow your clinician's guidance on whether to continue, switch, or return the product.
  4. If you have experienced a reaction, seek medical care and report it to the FDA via MedWatch.
  5. For the official FDA notice, visit the FDA recalls page.

Consumer Contact (per FDA)

Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-1789-2014.

Childrens Medical Ventures

FDA

About the U.S. Food and Drug Administration

The FDA regulates drugs, medical devices, food, cosmetics, and tobacco. Adverse event reports and recall notices are the main public safety signal.

Visit FDA.gov →

📣 Report an adverse event to the FDA (MedWatch)

If you had an adverse reaction, side effect, or device malfunction from this medication or device, report it to the FDA through MedWatch. Talk to your doctor or pharmacist first — do not stop a prescription on your own. Reports help the FDA detect rare or emerging safety signals and can lead to label changes or recalls.

How to report to FDAFile a report at FDA

Childrens Medical Ventures Recall FAQ

Childrens Medical Ventures is the subject of a medical devices safety report: Philips/Children's Medical Ventures Smart Monitor 2 Professional Series infant apnea monitor.. The notice was published on April 29, 2014 by the U.S. Food and Drug Administration (FDA). Approximately 112 units are potentially affected.