Skip to main content
High RiskFDAfda-Z-2390-2016STERILITY ISSUE

Stryker Orthopaedics Specialty Triathlon Tibial Alignment Handle with Secondary Lock, Non-Sterile, Used to complement or replace the s...

Units Affected
47
Recall Date
June 23, 2016
Issuing Agency
Hazard
Sterility Issue

FDA Recall Notice

Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-2390-2016.

It was reported that the secondary locking mechanism, and its corresponding components, disassociated from the Specialty Triathlon Tibial Alignment Handle with Secondary Lock Assembly during surgery. Upon further investigation, it was discovered that the weld, which was intended to hold the secondary locking mechanism together, did not meet the weld size specified on the engineering drawing.

Corrective Action (per FDA)

Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-2390-2016.

Recall terminated by FDA.

✅ What you should do

  1. Stop using the product if you own it.
  2. Check the model number, lot code, or sell-by date against the recall notice above.
  3. Contact Stryker Howmedica Osteonics Corp. or the retailer where you bought it for a refund, replacement, or repair.
  4. For the most current official instructions, visit the FDA recall page.
  5. If you've been hurt by this product, report the incident to FDA.

Consumer Contact (per FDA)

Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-2390-2016.

Stryker Howmedica Osteonics Corp.

FDA

About the U.S. Food and Drug Administration

The FDA regulates drugs, medical devices, food, cosmetics, and tobacco. Adverse event reports and recall notices are the main public safety signal.

Visit FDA.gov →

📣 Report a food, supplement, or cosmetic problem to the FDA

If you had a reaction, found contamination, or experienced a labeling problem with this product, report it to the FDA. The agency uses consumer reports to track emerging safety signals and trigger recalls.

How to report to FDAFile a report at FDA

Stryker Howmedica Osteonics Corp. Recall FAQ

Stryker Howmedica Osteonics Corp. is the subject of a medical implants safety report: Stryker Orthopaedics Specialty Triathlon Tibial Alignment Handle with Secondary Lock, Non-Sterile, Used to complement or replace the s.... The notice was published on June 23, 2016 by the U.S. Food and Drug Administration (FDA). Approximately 47 units are potentially affected.