Skip to main content
High RiskFDAfda-Z-2396-2021LACERATION

TRUMATCH CT PIN GUIDE FEM R- TruMatch Patient Specific Instruments are intended to be used as patient-specific surgical instrumentation t...

Units Affected
2
Recall Date
August 10, 2021
Issuing Agency
Hazard
Laceration

FDA Recall Notice

Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-2396-2021.

Software: A coding error associated with the Fast3D Segmentation software. During the scanning process the three images required of the hip, knee and ankle may not be aligned, causing the anatomic landmark locations to potentially result in limb malalignment.

Corrective Action (per FDA)

Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-2396-2021.

Recall terminated by FDA.

🩺 Talk to your doctor or pharmacist

Product Recall Tracker does not provide medical advice. Any specific instructions about this medication or device come directly from the FDA notice above. Before making any change to your treatment:

  1. Do not stop, change, or discard your prescription on your own.
  2. Call your doctor or pharmacist and ask whether this FDA notice applies to your prescription (have the brand name, dosage, and lot number ready if you can find them).
  3. Follow your clinician's guidance on whether to continue, switch, or return the product.
  4. If you have experienced a reaction, seek medical care and report it to the FDA via MedWatch.
  5. For the official FDA notice, visit the FDA recalls page.

Consumer Contact (per FDA)

Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-2396-2021.

DePuy Orthopaedics, Inc.

FDA

About the U.S. Food and Drug Administration

The FDA regulates drugs, medical devices, food, cosmetics, and tobacco. Adverse event reports and recall notices are the main public safety signal.

Visit FDA.gov →

📣 Report an adverse event to the FDA (MedWatch)

If you had an adverse reaction, side effect, or device malfunction from this medication or device, report it to the FDA through MedWatch. Talk to your doctor or pharmacist first — do not stop a prescription on your own. Reports help the FDA detect rare or emerging safety signals and can lead to label changes or recalls.

How to report to FDAFile a report at FDA

Depuy Orthopaedics, Inc. Recall FAQ

Depuy Orthopaedics, Inc. is the subject of a surgical devices safety report: TRUMATCH CT PIN GUIDE FEM R- TruMatch Patient Specific Instruments are intended to be used as patient-specific surgical instrumentation t.... The notice was published on August 10, 2021 by the U.S. Food and Drug Administration (FDA). Approximately 2 units are potentially affected.