Percepta Quad CRT-P MRI SureScan, REF W4TR04 Product Usage: The CRT-P models are implantable pulse generators with cardiac resynchron...
FDA Recall Notice
Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-2415-2018.
Potential for a device reset to occur in Percepta CRT-P MRI SureScan and Percepta Quad CRT-P MRI SureScan due to a timing interaction between the EffectivCRT Diagnostic and the Ventricular Safety Pacing feature (VSP).
Corrective Action (per FDA)
Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-2415-2018.
Recall terminated by FDA.
✅ What you should do
- Stop using the product if you own it.
- Check the model number, lot code, or sell-by date against the recall notice above.
- Contact Medtronic Inc., Cardiac Rhythm And Heart Failure (crhf) or the retailer where you bought it for a refund, replacement, or repair.
- For the most current official instructions, visit the FDA recall page.
- If you've been hurt by this product, report the incident to FDA.
Consumer Contact (per FDA)
Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-2415-2018.
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
About the U.S. Food and Drug Administration
The FDA regulates drugs, medical devices, food, cosmetics, and tobacco. Adverse event reports and recall notices are the main public safety signal.
Visit FDA.gov →📣 Report a food, supplement, or cosmetic problem to the FDA
If you had a reaction, found contamination, or experienced a labeling problem with this product, report it to the FDA. The agency uses consumer reports to track emerging safety signals and trigger recalls.
Medtronic Inc., Cardiac Rhythm And Heart Failure (crhf) Recall FAQ
Medtronic Inc., Cardiac Rhythm And Heart Failure (crhf) is the subject of a medical implants safety report: Percepta Quad CRT-P MRI SureScan, REF W4TR04 Product Usage: The CRT-P models are implantable pulse generators with cardiac resynchron.... The notice was published on June 21, 2018 by the U.S. Food and Drug Administration (FDA). Approximately 3,560 units are potentially affected.