Skip to main content
High RiskFDAfda-Z-2485-2012OTHER

Optilock 2.7 mm Screw Inserter. Intended Use: The OptiLock Upper Extremity Plating System is intended for fixation of fractures and os...

Units Affected
268
Recall Date
August 23, 2012
Issuing Agency
Hazard
Other

FDA Recall Notice

Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-2485-2012.

Field complaints indicate that recent shipments of the 14-400795 and 36505 drivers do not properly mate ("lock") into some of the specified screw heads. Drivers near MMC (Maximum Material Condition) or that are misshapen may not properly mate with screws at MMC. Design dimensions and tolerances of the drivers are not properly specified to prevent an incorrect fit between the drivers and the screw

Corrective Action (per FDA)

Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-2485-2012.

Recall terminated by FDA.

🩺 Talk to your doctor or pharmacist

Product Recall Tracker does not provide medical advice. Any specific instructions about this medication or device come directly from the FDA notice above. Before making any change to your treatment:

  1. Do not stop, change, or discard your prescription on your own.
  2. Call your doctor or pharmacist and ask whether this FDA notice applies to your prescription (have the brand name, dosage, and lot number ready if you can find them).
  3. Follow your clinician's guidance on whether to continue, switch, or return the product.
  4. If you have experienced a reaction, seek medical care and report it to the FDA via MedWatch.
  5. For the official FDA notice, visit the FDA recalls page.

Consumer Contact (per FDA)

Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-2485-2012.

Biomet, Inc.

FDA

About the U.S. Food and Drug Administration

The FDA regulates drugs, medical devices, food, cosmetics, and tobacco. Adverse event reports and recall notices are the main public safety signal.

Visit FDA.gov →

📣 Report an adverse event to the FDA (MedWatch)

If you had an adverse reaction, side effect, or device malfunction from this medication or device, report it to the FDA through MedWatch. Talk to your doctor or pharmacist first — do not stop a prescription on your own. Reports help the FDA detect rare or emerging safety signals and can lead to label changes or recalls.

How to report to FDAFile a report at FDA

Biomet, Inc. Recall FAQ

Biomet, Inc. is the subject of a medical devices safety report: Optilock 2.7 mm Screw Inserter. Intended Use: The OptiLock Upper Extremity Plating System is intended for fixation of fractures and os.... The notice was published on August 23, 2012 by the U.S. Food and Drug Administration (FDA). Approximately 268 units are potentially affected.