ClotTriever XL Catheter, 30 mm, REF: 41-102, used with Instructions for Use ClotTriever XL Catheter ClotTriever Thrombectomy System, Re...
⚠ Critical FDA Safety Notice — Talk to Your Doctor
The FDA has flagged this as a serious safety concern (laceration). Do not stop, change, or discard a prescribed medication or medical device on your own. Contact your doctor or pharmacist immediately to ask whether this notice applies to your prescription and what to do next. Any specific guidance below comes directly from the FDA — Product Recall Tracker does not provide medical advice.
FDA Recall Notice
Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-2587-2024.
Thrombectomy system catheter instructions for use updated to reduce adverse events due to the following circumstances: 1) Fibrous, firmly adherent, calcified material, or suspected tumor thrombus overloading catheter bag 2) Catheter pulled through internal jugular vein access site or not pulled away from heart; may lead to device entrapment, pulmonary embolism, venous cutdown, abdominal surgery.
Corrective Action (per FDA)
Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-2587-2024.
Recall ongoing. Follow firm instructions.
🩺 Talk to your doctor or pharmacist
Product Recall Tracker does not provide medical advice. Any specific instructions about this medication or device come directly from the FDA notice above. Before making any change to your treatment:
- Do not stop, change, or discard your prescription on your own.
- Call your doctor or pharmacist and ask whether this FDA notice applies to your prescription (have the brand name, dosage, and lot number ready if you can find them).
- Follow your clinician's guidance on whether to continue, switch, or return the product.
- If you have experienced a reaction, seek medical care and report it to the FDA via MedWatch.
- For the official FDA notice, visit the FDA recalls page.
Consumer Contact (per FDA)
Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-2587-2024.
Inari Medical - Oak Canyon
About the U.S. Food and Drug Administration
The FDA regulates drugs, medical devices, food, cosmetics, and tobacco. Adverse event reports and recall notices are the main public safety signal.
Visit FDA.gov →📣 Report an adverse event to the FDA (MedWatch)
If you had an adverse reaction, side effect, or device malfunction from this medication or device, report it to the FDA through MedWatch. Talk to your doctor or pharmacist first — do not stop a prescription on your own. Reports help the FDA detect rare or emerging safety signals and can lead to label changes or recalls.
Inari Medical - Oak Canyon Recall FAQ
Inari Medical - Oak Canyon is the subject of a surgical devices safety report: ClotTriever XL Catheter, 30 mm, REF: 41-102, used with Instructions for Use ClotTriever XL Catheter ClotTriever Thrombectomy System, Re.... The notice was published on July 19, 2024 by the U.S. Food and Drug Administration (FDA). Approximately 2,488 units are potentially affected.