Skip to main content
CriticalFDAfda-Z-2695-2016STERILITY ISSUE

ADVANCE(R) II MODULAR TIBIAL BASE W/SCREWHOLES, Knee Tibial Base, REF KTTIHA20, SIZE 2 STD; SURFACE POROUS, HA COATED; FIT CEMENTLESS; U...

⚠ Critical FDA Safety Notice — Talk to Your Doctor

The FDA has flagged this as a serious safety concern (sterility issue). Do not stop, change, or discard a prescribed medication or medical device on your own. Contact your doctor or pharmacist immediately to ask whether this notice applies to your prescription and what to do next. Any specific guidance below comes directly from the FDA — Product Recall Tracker does not provide medical advice.

Units Affected
110
Recall Date
June 21, 2016
Issuing Agency
Hazard
Sterility Issue

FDA Recall Notice

Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-2695-2016.

ADVANCE(R) HA Coated tibial bases are being recalled due to a higher than expected revision rate as a result of component loosening.

Corrective Action (per FDA)

Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-2695-2016.

Recall terminated by FDA.

🩺 Talk to your doctor or pharmacist

Product Recall Tracker does not provide medical advice. Any specific instructions about this medication or device come directly from the FDA notice above. Before making any change to your treatment:

  1. Do not stop, change, or discard your prescription on your own.
  2. Call your doctor or pharmacist and ask whether this FDA notice applies to your prescription (have the brand name, dosage, and lot number ready if you can find them).
  3. Follow your clinician's guidance on whether to continue, switch, or return the product.
  4. If you have experienced a reaction, seek medical care and report it to the FDA via MedWatch.
  5. For the official FDA notice, visit the FDA recalls page.

Consumer Contact (per FDA)

Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-2695-2016.

MicroPort Orthopedics Inc.

FDA

About the U.S. Food and Drug Administration

The FDA regulates drugs, medical devices, food, cosmetics, and tobacco. Adverse event reports and recall notices are the main public safety signal.

Visit FDA.gov →

📣 Report an adverse event to the FDA (MedWatch)

If you had an adverse reaction, side effect, or device malfunction from this medication or device, report it to the FDA through MedWatch. Talk to your doctor or pharmacist first — do not stop a prescription on your own. Reports help the FDA detect rare or emerging safety signals and can lead to label changes or recalls.

How to report to FDAFile a report at FDA

Microport Orthopedics Inc. Recall FAQ

Microport Orthopedics Inc. is the subject of a medical devices safety report: ADVANCE(R) II MODULAR TIBIAL BASE W/SCREWHOLES, Knee Tibial Base, REF KTTIHA20, SIZE 2 STD; SURFACE POROUS, HA COATED; FIT CEMENTLESS; U.... The notice was published on June 21, 2016 by the U.S. Food and Drug Administration (FDA). Approximately 110 units are potentially affected.