Skip to main content
CriticalFDAfda-Z-1497-2019FIRE OR BURN

Alinity i Processing Module, 03R65 01, with the below components. a. TED (Thermo Electric Device) Engine, Reagent cooler; b. CAB...

⚠ Critical FDA Safety Notice — Talk to Your Doctor

The FDA has flagged this as a serious safety concern (fire or burn). Do not stop, change, or discard a prescribed medication or medical device on your own. Contact your doctor or pharmacist immediately to ask whether this notice applies to your prescription and what to do next. Any specific guidance below comes directly from the FDA — Product Recall Tracker does not provide medical advice.

Units Affected
373
Recall Date
April 5, 2019
Issuing Agency
Hazard
Fire Or Burn

FDA Recall Notice

Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-1497-2019.

Potential loose cable connections on the reagent cooler, which could result in temperature errors and eventual reagent cooler failure. Could cause burn and shock injuries to users and a delay in test results. Device design has mitigated this risk, but it is a possibility.

Corrective Action (per FDA)

Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-1497-2019.

Recall terminated by FDA.

🩺 Talk to your doctor or pharmacist

Product Recall Tracker does not provide medical advice. Any specific instructions about this medication or device come directly from the FDA notice above. Before making any change to your treatment:

  1. Do not stop, change, or discard your prescription on your own.
  2. Call your doctor or pharmacist and ask whether this FDA notice applies to your prescription (have the brand name, dosage, and lot number ready if you can find them).
  3. Follow your clinician's guidance on whether to continue, switch, or return the product.
  4. If you have experienced a reaction, seek medical care and report it to the FDA via MedWatch.
  5. For the official FDA notice, visit the FDA recalls page.

Consumer Contact (per FDA)

Source: U.S. Food and Drug Administration (FDA) — official FDA notice for recall FDA-Z-1497-2019.

Abbott Gmbh & Co. KG

FDA

About the U.S. Food and Drug Administration

The FDA regulates drugs, medical devices, food, cosmetics, and tobacco. Adverse event reports and recall notices are the main public safety signal.

Visit FDA.gov →

📣 Report an adverse event to the FDA (MedWatch)

If you had an adverse reaction, side effect, or device malfunction from this medication or device, report it to the FDA through MedWatch. Talk to your doctor or pharmacist first — do not stop a prescription on your own. Reports help the FDA detect rare or emerging safety signals and can lead to label changes or recalls.

How to report to FDAFile a report at FDA

Abbott Gmbh & Co. Kg Recall FAQ

Abbott Gmbh & Co. Kg is the subject of a diagnostic devices safety report: Alinity i Processing Module, 03R65 01, with the below components. a. TED (Thermo Electric Device) Engine, Reagent cooler; b. CAB.... The notice was published on April 5, 2019 by the U.S. Food and Drug Administration (FDA). Approximately 373 units are potentially affected.